Otto Bock Bionic Pro Instructions For Use Manual page 38

Table of Contents

Advertisement

Use
FAULT
Not all sensors are displayed in
system information
Sensor is not being charged
User interface working at slower
pace
38
PROBABLE CAUSE
Sensor switched off
Identified by:
Sensor not illuminated
Sensor in sleep mode
Identified by:
Sensor not illuminated
Sensor battery drained
Identified by:
Sensor switched on and not illu­
minated
WLAN connection to sensor lost
Identified by:
Sensor no longer illuminated
blue
Sensor defective
Identified by:
Sensor not illuminated despite
previous steps
USB port dirty
Identified by:
Dirt in USB port
USB port defective
Identified by:
Sensor not charged despite pre­
vious steps
Base station not switched on
Interference due to other HF
devices
SUGGESTED REMEDY
Switch sensor on
Shake
sensor
("Charging
sensors in the base station and
switching them on prior to meas­
uring" section – see page 15)
Charge sensor
Move sensor into range of router
Switch sensor off and on again
Contact service
Carefully clean USB port with
compressed air (at 20 cm dis­
tance) or a vacuum cleaner
Contact service
Switch base station on
Refer to the instructions in the
"Preventing
electromagnetic
interference"
section
page 17)
743A250 Bionic Pro
(see

Advertisement

Table of Contents
loading

This manual is also suitable for:

743a250743a250 bionic pro