5.7.1.3 "New firmware for sensor" system message

49
Take the following steps for the "New firmware for sensor" error message:
► Insert all switched on sensors in the base station.
► Tap the "Update now"   and "OK"   buttons.
► Wait until the sensor firmware has been updated.

5.7.2 Fault table

FAULT
System has no Internet
Browser notification: "Page can­
not be opened"
Tablet screen remains dark
743A250 Bionic Pro
PROBABLE CAUSE
Network cable not connected to
router
Identified by:
Right LED on base station not
illuminated
Company network has no Inter­
net
Identified by:
Icon on router screen
No WLAN connection
Identified by:
No WLAN icon can be seen in
menu bar
Router is off
Identified by:
LEDs on sensors flashing rapidly
(white)
Centre LED on base station
flashing rapidly
iPad connected to wrong WLAN
Identified by:
Tablet is not connected to Bionic
Pro router in WLAN settings
Base station switched off
Identified by:
LEDs on base station not lit
Base station in configuration
mode (button on back accident­
ally pressed)
Identified by:
Centre LED (WLAN) on base
station flashing slowly
Battery discharged
Identified by:
Tablet cannot be switched on
SUGGESTED REMEDY
Plug network cable into WAN
socket on the router
Check company network
Contact administrator
Provide
LAN
socket
proxy settings
Connect iPad to WLAN
Switch on WLAN on iPad
Connect router to mains network
Connect iPad to the WLAN of
the Bionic Pro router
Connect base station to mains
network
Unplug base station from mains
network for at least 3 seconds
and then plug back in again
Charge tablet
Use
without
37