5 Use

► NOTICE! In order to avoid damage use the product exclusively according to the instructions for use, the quick
reference guide and the online help of the product.
INFORMATION
The QR codes, nameplates and passwords contained in the illustrations are only samples and may not be used.
Only the QR codes in the iPad user manual are intended for use (may be changed by Apple).

5.1 Switching on

14
► Follow the sequence shown in the illustrations when switching on the product and adhere to the waiting times
so the product components connect with one another properly via WLAN.

5.2 Preventing electromagnetic interference

The individual product components (base station, sensors, router) are connected with each other via WLAN and
enable the required exchange of data. Other devices that produce electromagnetic fields or networks can cause
electromagnetic interference that slows the exchange of data or makes it impossible. This may mean that a meas­
urement needs to be repeated. Electromagnetic interference may be indicated by error messages on the user inter­
face or manifest itself by the user interface working more slowly. The following instructions help to prevent or elim­
inate electromagnetic interference.
► Do not use the product in areas with strong electromagnetic fields (e.g. in the proximity of magnetic resonance
imaging equipment, microwaves and machines).
► During use of the product, ensure that other high-frequency devices, such as mobile phones, smartphones,
tablets and routers do not come within the required safety distance of at least 300 mm to the product compon­
ents (base station, sensors, router).
► Refer to the instructions in the "Troubleshooting" (see page 35) section and its subsections in the event of
faults or error messages.
→ If this does not remedy the faults, switch off all other high-frequency devices.
→ If this does not remedy the faults, use the product in a different room.
→ If this does not remedy the faults, contact the manufacturer.