► Beim Anbringen der Sensoren außer den Warnhinweisen und der Abbildung in diesem Kapitel auch die fol­
genden Anweisungen beachten:
► In addition to the warnings and the illustration in this section, also observe the following instructions when
attaching the sensors:
: NOTICE! Start by attaching all the sensor holders to the patient and then 
er sensor holder.
► NOTICE! The sensor holders must fit as closely to the body as possible so they do not slip during movement,
as this would otherwise distort the measurement. Touching and impacts also have to be avoided for the same
reason.
► CAUTION! The warnings at the beginning of this section must be observed when positioning the sensor hold­
ers on the skin.
► NOTICE! When the sensor holders are not positioned on the skin, tight-fitting clothes (e.g. cycling shorts) or
thin single-use fabric as an intermediate layer have to be used.
► NOTICE! To apply the two sensor holders FL + FR, first take off the shoes and slide the sensor holders over
the feet. Put the shoes on again and position the sensor holders in the middle of the hindfoot and slide the
plastic tongue on the front underneath the shoelaces. The sensor must sit above the sole of the shoe.
: When lengthening the straps, make sure that the hook-and-loop sections 
the same direction.
: For prostheses without a foam cover, insert the sensor into the sensor holder that is attached to the tube
adapter at the level of the calf.
: If the orthosis or prosthesis prevents a sensor from being attached in the ideal location, attach it as closely
as possible to the ideal location.
: Use all seven sensors for the 10 metre gait analysis. Position the TL and CL sensors laterally on the left leg.
: Do not use the TL and CL sensors for the drop foot fitting assessment.
► Observe the following notices for correct positioning of the sensors:
P (pelvis) 
Sensor holder with straight back
Positioning: in the centre of the lumbar spine region as close as possible to the body
T (thigh), TL (thigh, left):  thigh
Sensor holder with slightly rounded back
Positioning: lateral in the centre of the upper third of the thigh
C (calf), CL (calf, left): calf
Smaller sensor holder with pronounced curve on back
Positioning: lateral at the position with the largest circumference; thigh and calf sensor aligned on the same
axis
FL (foot left), FR (foot right): left foot, right foot
Positioning: sensor holder centred on the hindfoot right over the sole of the shoe; plastic tongue under the
shoelaces
743A250 Bionic Pro
insert each sensor in the prop­
of both straps are aligned in
Use
23