12.3 Glossary
The term "Axon" stands for Adaptive exchange of neuroplacement data. The Axon-Bus is an
Ottobock innovation for the field of exoprosthetics: a data transmission system, derived from
safety-related bus systems in the aviation and automobile industries. For the user this means
enhanced safety and reliability because of a considerably reduced sensitivity to electromagnetic
interference in comparison with conventional systems.
12.4 Directives and manufacturer's declaration
12.4.1 Electromagnetic environment
This product is designed for operation in the following electromagnetic environments:
Operation in a professional healthcare facility (e.g. hospital, etc.)
Operation in areas of home healthcare (e.g. use at home, use outdoors)
Observe the safety notices in the section "Information on proximity to certain areas" (see
page 23).
Electromagnetic emissions
Interference meas­
urements
HF emissions accord­
ing to CISPR 11
Harmonics according
to IEC 61000-3-2
Voltage fluctu­
ations/flicker accord­
ing to IEC 61000-3-3
Electromagnetic interference immunity
Phenomenon
Electrostatic dis­
charge
High-frequency elec­
tromagnetic fields
Magnetic fields with
rated power frequen­
cies
Electrical fast transi­
ents/bursts
Surges
Line against line
Conducted interfer­
ence induced by high-
frequency fields
Compliance
Group 1/class B
Not applicable –
power below 75 W
Product meets the
requirements of the
standard.
EMC basic standard
or
test procedure
IEC 61000-4-2
IEC 61000-4-3
IEC 61000-4-8
IEC 61000-4-4
IEC 61000-4-5
IEC 61000-4-6
Electromagnetic environment directive
The product uses HF energy exclusively for its
internal functioning. Its HF emissions are
therefore very low, and interference with neigh­
bouring electronic devices is unlikely.
Interference immunity test level
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air,
10 V/m
80 MHz to 2.7 GHz
80% AM at 1 kHz
30 A/m
50 Hz or 60 Hz
± 2 kV
100 kHz repetition rate
± 0.5 kV, ± 1 kV
3 V
0.15 MHz to 80 MHz
6 V in ISM and amateur frequency bands
between 0.15 MHz and 80 MHz
80% AM at 1 kHz
35