Error while charging the product
Battery charger
12.2.3 Status signals
Battery charger is connected
Battery charger
Battery charge level
Charging
Event
receptacle
Battery is fully charged
Battery charge level is 100%
Battery charge level is over 50%
Battery charge level is less than 50%
34
Explanation/correction
Axon-Bus prosthetic system is not charging
•
Disconnect and then reconnect the charging plug.
•
Clean charger contacts.
Charger is overheated or overcooled
•
Disconnect charging plug and allow charger to cool
down/warm up.
Battery charger or power supply unit is defective
•
Consult prosthetist
Explanation
Power supply and battery charger operational
Battery is charging
Battery is charging, battery 25% charged
Battery is charging, battery 50% charged
Battery is charging, battery 75% charged
Battery is fully charged, battery 100% charged