Spare Channel Replacement Procedure - Agilent Technologies 86060A User Manual

Lightwave switches
Table of Contents

Advertisement

C A U T I O N
1 Disconnect the power cord from the instrument.
2 Remove the top cover using a #15 or #20 TORX driver.
3 Locate out the spare cable/connector. Refer to
4 Replace the defective port cable/connector assembly with the spare assembly.
5 Replace the top cover.
6 Connect the power cord to the instrument. Turn the instrument on.
7 Press Help, then enter the model number of the lightwave switch from the
8 Press the 4 key.
9 Perform this step only if the lightwave switch has multiple layers. Determine if
Servicing

Spare Channel Replacement Procedure

Spare Channel Replacement Procedure
This procedure is included so the instrument can be repaired quickly in the
field. Performing this procedure violates the calibration seal. When spare fibers
have replaced bad port fibers, the unit should be sent to an Agilent
Technologies Service Center, at your earliest convenience, to have new port
fiber/connectors spliced into the instrument and to have the instrument
calibrated.
Figure 5-2 on page
5-6. The spare cable/connector is on the bottom of the
switch assembly, secured with a tie-wrap. Remove the tie-wrap.
Switches with 50 channels or less have one spare fiber. Switches with more
than 50 channels have two spare fibers, labeled S1 and S2.
Note the label on the spare cable/connector and record the identification
number (S1 or S2) for use later on in this procedure.
Identification number: . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ____________
keypad: 86060, 86061, or 86062.
the lightwave switch has multiple switch layers installed. (A switch layer is a
switch matrix of "A" ports and "B" ports. Normal lightwave switches have only
one layer installed.) If the rear panel shows more than one set of "A" ports and
"B" ports, the instrument has multiple switch layers. The display will prompt