Lauda PROLINE Edition X P 5 C Operating Instructions Manual page 65

Hide thumbs Also See for PROLINE Edition X P 5 C:
Table of Contents

Advertisement

Proline Edition X heating thermostats
 Using the cursor keys move the black background to the field which you would like to change.
It can be edited by pressing the key
 The soft key
taken from the previous segment with the exception of the tolerance field. The tolerance is al-
ways specified as 0.00. All following segment lines will be moved one line downwards.
 In the above window Segment 1 was created in this way.
 Continue with
 If there is no entry in the "Time" field, the bath temperature is approached as quickly as possi-
ble. With a time entry the final temperature is obtained exactly after the time expires (ramp).
 The entry in the field "Tolerance" field defines how accurately the final temperature is to be ob-
tained before the next segment is processed. In case there is no additional cooling, you should
select a more generous tolerance limit. Check and observe the transient effect using the
"graphic display" ( Section 7.9).
If the tolerance range has been selected too small, it may be that the program does not
continue, because the required tolerance is never achieved.
External temperature control: Especially with ramps, a too close tolerance range can
cause undesired delays in the start phase of the ramp.
No.
Pump
Start
-------
1
Pump
Menu
YACE0099 / 10/9/2018
Insert inserts in the marked line a new segment which has a default value
to the fields  "Time"  "Tolerance". See program example in ( 7.10.1).
Out 1
Out 2
-------
-------
4
-------
-------
End
Insert
(see following pages).
 Then continue with
signal output setting.
Out 3
 The right-hand part of the entry table ap-
-------
pears as shown on the left.
-------
 Here, in the "Pump" field, the pump level
and, in the fields "Out 1" to "Out 3", the
contact outputs of the contact mode (ac-
cessory) can be programmed. With the
setting "-------" the starting value is re-
tained which was either set before the
program start or was defined by a previ-
ous segment in the running program.
Further details are given on the following
pages.
Delete
Starting up
to the pump and
65

Advertisement

Table of Contents
loading

Table of Contents