Proline Edition X heating thermostats
throughput through the external load as large as possible.
2.2. Xpf Setting:
when oscillating with short period occur (i.e. 30 s) Xpf lower, otherwise higher,
in case of bad thermal coupling and large thermal mass high (i.e. 2 — 5, or even higher),
in case of good thermal coupling and small thermal mass small (i.e. 0.2 — 0.7),
when rapid temperature response is required simple internal control should be preferred. Otherwise
select very small Xpf (0.05 — 0.1).
3.
Setting the master controller (PIDT
Start with setting Auto and proceed with Manual only when necessary.
3.1. Kpe setting:
Start with empirical values from table ( 0).
In case of oscillations (with large period, i.e.10 min) Kpe higher, otherwise lower.
3.2. – Tne/ Tve/ Tde setting:
Start with empirical values from table ( 0) and with high numbers (Tne = 70 s — 200 s;
Tve = 50 s — 150 s),
with lower numbers faster approach, otherwise slower approach with lower oscillations,
Tve: to reduce overshot Tve higher, otherwise lower,
Tde (damping for Tve): in general approximately 10 % of Tve.
4.
Correction limitation ( or outflow temperature limitation) ( 7.13.2) and temperature limits (Til/ Tih)
( 7.8.3)
make settings in accordance with the boundary conditions. Examples:
Heat transfer liquid
Water
Ethanol
Tools to watch the time behavior:
Graph mode of the Command remote control,
LAUDA Wintherm PC-program.
YACE0099 / 10/9/2018
-controller):
1
Correction limitation
depending on the external vessel size
and the heat transfer liquid
Starting up
Til
Tih
2 °C
95 °C
Minimum
40 °C
77