The bath level is acquired by the SelfCheck Assistant in 8 stages. A permanent display is provided only
with the Command remote control. If the minimum level is undercut, the pump and heater are switched
off. The reaction of the thermostat in case of overfill can be set to simply display a warning, to display a
warning and switch off the heater or to switch off the unit completely with pump and heater.
When the level is too low, with overtemperature, or with other alarms the SelfCheck Assistant switches
the heater off on all poles. The pump is also switched off.
This switch-off under fault conditions is retained, i.e. after the fault is rectified, the fault must be reset (re-
leased) on the Master operating panel with the
Other unit functions are described in the appropriate sections and in Section 7. (Starting up).
4.6

Programmer and ramp function

Master Version:
No programmer provided.
Command remote control:
The units are equipped with a programmer function, which enables five temperature/ time programs to
be saved. Each program consists of a number of temperature/ time segments. These also include de-
tails of how often the program is to be executed. Up to 150 segments can be distributed amongst the
five programs.
With the ramp function, a rate of change can be directly entered in °C per unit time. ( 7.10).
4.7

Interfaces

Master Version:
The Master unit is equipped with the following sockets at the back of the control head:
For the connection of an external Pt100 temperature sensor (10S).
Two sockets (70S) for the connection of components via the LAUDA equipment bus (cooling sec-
tion, Command remote control, external solenoid valve, etc.).
Command remote control:
The Command remote control is equipped as standard with the following sockets:
For the connection of an external Pt100 temperature probe (10S).
Two sockets (70S) for the connection of components via the LAUDA equipment bus (cooling sec-
tion, Command remote control, external solenoid valve, etc.)
One RS-232/485 interface (65S) at the back of the Command remote control.
YACE0099 / 10/9/2018
key.
Device description
19