Lauda PROLINE Edition X P 5 C Operating Instructions Manual page 94

Hide thumbs Also See for PROLINE Edition X P 5 C:
Table of Contents

Advertisement

Command
IN_PAR_06
IN_PAR_07
IN_PAR_09
IN_PAR_10
IN_PAR_14
IN_PAR_15
IN_DI_01
IN_DI_02
IN_DI_03
IN_DO_01
IN_DO_02
IN_DO_03
IN_MODE_00
IN_MODE_01
IN_MODE_02
IN_MODE_03
IN_MODE_04
TYPE
VERSION_R
VERSION_S
VERSION_B
VERSION_T
VERSION_A
VERSION_V
VERSION_D
VERSION_M_0
VERSION_M_1
VERSION_M_3
VERSION_M_4
VERSION_M_5
STATUS
STAT
RMP_IN_00_XXX
RMP_IN_01
RMP_IN_02
94
Proline Edition X heating thermostats
Read control parameter TvE (0 = OFF)
Read control parameter TdE
Read value of correction limitation
Read control parameter XpF
Read setpoint offset
Read control parameter PropE
State of contact input 1: 0 = open/ 1 = closed
State of contact input 2: 0 = open/ 1 = closed
State of contact input 3: 0 = open/ 1 = closed
State of Contact output 1:
0 = make-contact open/ 1 = make-contact closed
State of Contact output 2:
0 = make-contact open/ 1 = make-contact closed
State of Contact output 3:
0 = make-contact open/ 1 = make-contact closed
Keyboard Master: 0 = free / 1 = locked
Control: 0 = int. / 1 = ext. Pt100 / 2 = ext. Analogue / 3 = ext. Serial
Standby operation: 0 = Device ON / 1 = Device OFF
Keyboard Command Remote Control: 0 = free / 1 = locked
Setpoint offset source: 0 = normal/ 1 = ext. Pt/ 2 = ext. analogue/ 3 =
ext. serial
Read device type (response e.g. "P 8")
Read software type of control system
Read software type of protection system
Read software type of Command
Read software type of cooling system
Read software type of analogue module
Read software type of RS-232/485 module
Read software type of digital module
Read software type of solenoid valve (cooling water)
Read software type of solenoid valve (automatic refilling)
Read software type of solenoid valve (shut-off valve 1)
Read software type of solenoid valve (shut-off valve 2)
Read software type of high temperature cooler
Read equipment status 0 = OK, -1 = error
Read error diagnosis response:
XXXXXXX  X = 0 no error, X = 1 error
1. char = fault
2. char = alarm
3. char = warning
4. char = over temperature
5. char = low bath level
6. char = high bath level (at adjustment alarm)
7. char = no external control variable
Read a programme segment XXX
(response: e. g. 030.00_010.00_005.00_001.00 > set point tempera-
ture 30.00 °C, time = 10 min, tolerance = 5.00 K, pump level = 1)
Read the current segment number
Read the set number of programme runs
Interface modules
Explanation
/
YACE0099
10/9/2018