Manual cleaning/disinfection

Keep working ends open for cleaning.
When cleaning instruments with movable hinges,
ensure that these are in an open position and, if
applicable, move the joint while cleaning.
After
manual
cleaning/disinfection,
accessible surfaces visually for residues.
Where necessary, repeat the cleaning process.
Manual cleaning with immersion disinfection
Phase
Step
I
Cleaning
II
Intermediate
rinse
III
Disinfection
IV
Final rinse
V
Drying
D–W:
Drinking water
DM-W:
Demineralized water
RT:
Room temperature
Phase I
Completely immerse the product in the cleaner/
disinfectant solution. Ensure that all accessible
surfaces are moistened.
Clean the product under running tap water with a
suitable cleaning brush where necessary for as long
as it takes to remove all discernable residues.
check
T
t
Conc.
[°C/°F]
[min]
[%]
RT
15
2
(cold)
RT
1
-
(cold)
RT
15
2
(cold)
RT
1
-
(cold)
RT
-
-
Water
quality
D–W
BBraun Stabimed; phenol aldehyde-
free and QAV-free
D–W
-
D–W
BBraun Stabimed; phenol aldehyde-
free and QAV-free
DM–W
-
-
-
For products with concealed crevices, lumens or
complex geometries, brush non-visible surfaces
with a suitable cleaning brush (brush length:
30 mm/∅: 4,5 mm, e.g. a TA011944) for at least
1 min or until all discernable residues have been
removed.
Maneuver
non-rigid
components,
adjustable screws, hinges etc during cleaning.
Chemical
such
as
7