Loading and Unloading the Test Cell

The OX-TRAN Model 2/28 uses a pneumatic system to clamp and unclamp the Test Cell. The Test cell
must be manually loaded and unloaded from the instrument.
To Unload the Test Cell from the Cell Tray, follow the instructions below (refer to Figure 5-3):
1.
Press the Cell Load/Unload button on the front of the instrument (see Figure 2-1).
2.
Open the Cell Tray completely by gently pulling straight back from the front panel.
3.
Grasp the Test Cell by the front and back edges and lift straight up.
To Load the Test Cell into the Cell tray, follow the instructions below (refer to Figure 5-3 and 5-4):
1.
Grasp the Test Cell by the front and back edges and lower it straight down.
2.
Close the Cell Tray completely by gently pushing straight towards the front panel.
3.
Keep your finger holding the tray at its closed position.
4.
While holding the Cell Tray, press the Cell Load/Unload button to clamp the cell (see Figure
5-4).
5-6

Figure 5-3: Loading the Test Cell

Revision J
OX-TRAN Model 2/28 Operator's Manual
MOCON, Inc.