X-Ray Tubes - AGFA DX-D 300 User Manual

U-arm
Hide thumbs Also See for DX-D 300:
Table of Contents

Advertisement

7.4

X-RAY TUBES

E7884X
E7252X
E7254X
E7869XX
0171F EN 20181009
Low Speed -- Rotating Anode, Focal Spots: 0.6 mm / 1.2 mm
Anode kHU / kVp: 300 kHU / 150 kVp, Target Angle: 12°
Maximum Specified Energy Input in 1 hour: 150 kVp @ 3408 mAs
Inherent Filtration of X-ray Source (Tube + Collimator): refer to Identification Label
High Speed -- Rotating Anode, Focal Spots: 0.6 mm / 1.2 mm
Anode kHU / kVp: 300 kHU / 150 kVp, Target Angle: 12°
Maximum Specified Energy Input in 1 hour: 150 kVp @ 5760 mAs
Inherent Filtration of X-ray Source (Tube + Collimator): refer to Identification Label
High Speed -- Rotating Anode, Focal Spots: 0.6 mm / 1.2 mm
Anode kHU / kVp: 400 kHU / 150 kVp, Target Angle: 12°
Maximum Specified Energy Input in 1 hour: 150 kVp @ 4800 mAs
Inherent Filtration of X-ray Source (Tube + Collimator): refer to Identification Label
High Speed -- Rotating Anode, Focal Spots: 0.6 mm / 1.2 mm
Anode kHU / kVp: 600 kHU / 150 kVp, Target Angle: 12°
Maximum Specified Energy Input in 1 hour: 150 kVp @ 5189 mAs
Inherent Filtration of X-ray Source (Tube + Collimator): refer to Identification Label
DX-D 300 U-Arm
User Manual
89

Advertisement

Table of Contents
loading

Table of Contents