4.6
AUTOMATIC COLLIMATOR
Transverse Field Size
Longitudinal Field Size
Source to Image Distance
0171F EN 20181009
Collimator controls:
1.
Collimator Display
2.
Manual Blade Knobs
3.
Automatic Mode Indicator (Green)
4.
System not Ready Indicator (Red)
5.
Manual Mode Indicator (Yellow)
6.
Change of Filter
7.
Collimator Lamp Control (Led "ON")
8.
Retractable Metric Tape
9.
Laser Pointer Window
10.
Laser Pointer On/Off Button
1
Manual Blade Knobs
When pressing the Collimator Lamp push-button, the Collimator light and an
optional Laser light turn on. They remain lighting for 30 seconds before they
switch Off automatically (lighting time can be configured).
Exposure field on the DR Detector is automatically adjusted, or when in manual
mode by setting the two knobs. The Collimator Display shows the selected field
size (cross / long).