SECTION 7
7.1
ENVIRONMENTAL CONDITIONS
ATMOSPHERIC PRESSURE (hPa)
MIN
700 hPa
500 hPa
.
Note
7.2
U-ARM
7.2.1
POWER LINE REQUIREMENTS
VOLTAGE
230 / 240 V~
Single Phase
7.2.2
INFORMATION RELATED TO RADIATION
0171F EN 20181009
TECHNICAL SPECIFICATIONS
RELATIVE HUMIDITY (%)
MAX
MIN
1060 hPa
30%
TRANSPORT & STORAGE
1060 hPa
10%
These environmental conditions do not include the Digital
Detector. Refer to the Digital Detector Documentation.
FREQUENCY
50/60 Hz
Radiation Output Accuracy: C.V. (Coefficient of Variation) ≤0.05
(Reproducibility related to loading factors)
Maximum Symmetrical Radiation Field:
Measured at 75 kVp: 160 mm in "X" axis and 240 mm in "Y" axis.
Measured at 125 kVp: 80 mm in "X" axis and 240 mm in "Y" axis.
(Test performed at a distance from the Focal Spot of 1200 mm, in accordance with IEC 60806: 1984)
MAX
WORKING
75%
90%
MINIMUM INPUT POWER
REQUIRED
2.5 kVA
DX-D 300 U-Arm
User Manual
AMBIENT TEMPERATURE
MIN
MAX
10 o C (50 o F)
35 o C (95 o F)
-20 o C (- -4 o F)
70 o C (158 o F)
POWER CONSUMPTION
IN STAND-BY
80 W
83