DX-D 300 U-Arm
User Manual
Table 4-2 (Cont.)
Information Messages
MESSAGE ON SCREEN
Bumper activated
Program Aborted: Invalid Parameters
Program Aborted: Incorrect button pressed
Stitching
Press Move button for Stitching Initial setup
Press Move button for Stitching Initial setup
Press Move button to extra image position
Detector Orientation
Tray Out
Tube Rotated and Detector inside
Emergency Stop
76
WHAT TO DO
Press the Safety Lock Button
Press the Safety Lock Button and try
again
Press "Move" button
Press "Move" button
Stitching is not possible with Detector in
landscape position. Turn the Detector to
Portrait position.
Insert the Detector Tray.
X-Rays inhibited until the Tube is
pointing the Detector. Turn the Tube
towards the Detector.
Message associated to a Fatal Error.
Turn the Unit OFF/ON, if error remains,
call field service.