AGFA DX-D 300 User Manual page 43

U-arm
Hide thumbs Also See for DX-D 300:
Table of Contents

Advertisement

GUIDANCE AND MANUFACTURER'S DECLARATION - ELECTROMAGNETIC IMMUNITY
The Radiographic Room DX-D300 is intended for use in an electromagnetic environment specified below.
The customer or Operator of thisRadiographic Room DX-D300 should assure that it is used in such an environment.
Immunity Test
Radiated RF EM fields
IEC 61000-4-3
Proximity fields from RF
Proximity fields from RF
wireless Communications
equipment
IEC 61000-4-3
Conducted disturbances
induced by RF fields
IEC 61000-4-6
NOTE - - The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz;
26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
The amateur radio bands between 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz; 3.5 MHz to 4.0 MHz; 5.3 MHz to 5.4 MHz; 7 MHz to 7.3 MHz; 10.1 MHz
to 10.15 MHz; 14 MHz to 14.2 MHz; 18.07 MHz to 18.17 MHz; 21.0 MHz to 21.4 MHz; 24.89 MHz to 24.99 MHz; 28.0 MHz to 29.7 MHz; and 50.0 MHz
to 54.0 MHz.
0171F EN 20181009
(IEC 60601-1-2:2014)
IEC 60601-1-2:2014
Test Level
3 Vrms
from 80 MHz to 2.7 GHz
(80% AM at 1 kHz)
Refer to next table "IMMUNITY
Refer to next table "IMMUNITY
REQUIREMENTS FOR RF
WIRELESS COMMUNICATIONS
EQUIPMENT"
3 Vrms
from 150 kHz to 80 Mhz
6 Vrms in ISM bands
from 150 kHz to 80 MHz
(80% AM at 1 kHz)
Compliance Level
3 Vrms
from 80 MHz to 2.7 GHz
(80% AM at 1 kHz)
Refer to next table "IMMUNITY
Refer to next table "IMMUNITY
REQUIREMENTS FOR RF
WIRELESS COMMUNICATIONS
EQUIPMENT"
3 Vrms
from 150 kHz to 80 Mhz
6 Vrms in ISM bands
from 150 kHz to 80 MHz
(80% AM at 1 kHz)
DX-D 300 U-Arm
User Manual
Electromagnetic environment -
guidance
Portable RF communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm to any part of the
equipment, including cables specified by
manufacturer. Otherwise, degradation of the
performance of this equipment could result.
performance of this equipment could result.
35

Advertisement

Table of Contents
loading

Table of Contents