Anticrushing System: When the U-Arm is moving down or when it rotates,
it will always reverse the movement if the Anticrushing System is activated to
avoid something kept between the Detector Assembly and the floor.
2
When the U-Arm is moving up and the Anticrushing System is activated, it will
make the reverse movement or stop depending on the following situations:
The system only reverses down automatically when the Anticrushing
System is activated and the angle is between --30
interval where the Detector Assembly could be under the table). In any
other case it stops.
o
- -30
2
3
In the rest of the angle ranges (from 51
position, the system just stops.
51
In case the Unit remains continuously blocked, it can be moved by
pressing and holding the "Move" button and then pressing the
movement button required.