AGFA DX-D 300 User Manual page 62

U-arm
Hide thumbs Also See for DX-D 300:
Table of Contents

Advertisement

DX-D 300 U-Arm
User Manual
+
Any movement
Button
54
MOVE BUTTON: Press and hold this button once a Program number is
selected in the Program Screen of the Touch Screen Console. The U-Arm
moves automatically to the selected position. In case the button is released
before reaching the selected position, a message appears on the Screen:
"Move button released. Press Move Button to continue". Once the Program
position is reached, a message appears on the Screen: "Position reached" and
the Safety Lock button is activated, the Operator must press it before
performing any other motion. If the operator modifies the Program position, the
screen displays: "Programmed position modified by the operator". Also refer to
Section 4.1.3.
MOVE + ANY MOVEMENT BUTTON: In case the Unit remains continuously
blocked (the Safety Lock button is activated), it can be moved by pressing and
holding the "Move" button and then pressing the movement button required.
ROTATION DISPLAY: This Display shows the angle performed by the U-Arm
and the angle performed by the Detector.
It also displays some error messages.
PROXIMITY SENSOR 1: The U-Arm is provided with a photocell light that
detects the patient on the Table, once this happens, Low Speed is automatically
activated. Also lowering the U-Arm below the Table-Top and rotation will not be
allowed.
PROXIMITY SENSOR 2: This second photocell disables all movements
(except for U-Arm up movement) when the photocell light is interrupted. Once
the beam is cleared all movements are enabled.
The Touch Screen shows a message when the Proximity Sensor is activated.
Proximity Sensor 2
Proximity Sensor 1
0171F EN 20181009