7. 3. 6. 2 Calibration 2 of 2: Power from RRH 1 RFHD 2 -> RRH 1 RFHD 1
1. Calibrate the Source (Tx) signal path
S9100A 5G Multi-Band Vector Transceiver, Startup Guide
This calibration is used to calibrate FR2 ports
RF Tx/Rx 2 ➁
(with
S9100A Option 022 mmWave Transceiver
with High IF and Blocker.
When performing this Power Calibration, the Source (Tx) output
power level is set to -5 dBm by default. The power level from
M9410A PXIe VXT #1 is calibrated while the RF Output for
M9410A PXIe VXT #2 is set to Off.
a. Select the Power Calibration tab.
b. Select the Operation drop-down arrow and select
c. Select the Calibration Step(s) drop-down arrow and select
d. Select the Port Selection drop-down arrow and
select
RRH 1 RFHD 2 -> RRH 1 RFHD 1
e. Select
Start the calibration
f. Connect the U8487A Power Sensor or equivalent to
M1740A mmWave Transceiver connector
g. Select
to continue the calibration.
Ok
View the
Results Source
Corrections are stored in:
Tx-RFIO_2-CHANNELPOWER.csv
used as the RF Output) on an
.
and follow the prompts.
RF Tx/Rx 2 ➁
tab when the calibration is complete.
S9100ID_<SerialNumber>-
7 Run Power Calibration
.
Calibration
.
Complete
.