7 Run Power Calibration
7. 3. 2 Using X-Apps to Perform Alignments
1. If it is not already running, start the Modular TRX interface:
2. Select the settings icon (top right corner of the display).
3. Select Alignments (all alignments must be performed before calibrations).
74
All alignments must be performed before Power Calibration.
The Modular TRX interface can be started as follows:
Select the Start menu (lower-left corner icon)
> Scroll down to "K" applications
> Select the Keysight Modular Transceiver drop-down arrow
> Scroll down the list and select LaunchModularTRX
a. Select Align Now All.
b. Select Align IF Cable.
S9100A 5G Multi-Band Vector Transceiver, Startup Guide