Using The Pressure Display As Diagnostic Tool - Agilent Technologies SD-1 User Manual

Isocratic solvent delivery module
Table of Contents

Advertisement

Using the Pressure Display as Diagnostic Tool

The sensitivity of the pressure display is within 68.9 kPa (10 psi). The
pressure display can be used as a diagnostic tool. Following issues cause
characteristic pressure fluctuations in HPLC systems (parameters valid for
HPLC operating at normal pressure):
• Zero to several hundred kPa: Bubbles in the solvent
• 68.9 – 275.8 kPa Sticking check valve
If fluctuations > 68.9 kPa try to eliminate bubbles in the solvent (see
H I N T
from the Liquid Head"
If this does not solve the problem, a check valve is probably sticking. It may be possible to
rectify this problem by cleaning the check valve (see
SD-1 Solvent Delivery - User Manual
Using the Pressure Display as Diagnostic Tool
on page 84).
Troubleshooting and Diagnostics
"Clearing Air Bubbles
"Cleaning Check Valves"
6
on page 95).
63

Advertisement

Table of Contents
loading

This manual is also suitable for:

G9302aG9303a

Table of Contents