Checking and/or Changing Power Fuses (F1)
When
If the Agilent SD-1 does not operate when the power cord is connected and the power switch is on,
replace the fuse(s). Fuses are located in the power module on the back of the Agilent SD-1.
Tools required
Description
Flat hat screwdriver
Parts required
p/n
R005100022
R005100029
Preparations
Check following probable causes for the power fail before starting exchanging the fuses:
• Power cord properly connected
• Power at the wall receptacle
Electrical shock
WARNING
➔
Wrong fuse
C A U T I O N
Damage to the module and loss of warranty.
➔
➔
1 Switch off the pump power and remove the AC power cord.
SD-1 Solvent Delivery - User Manual
Description
Fuse, 3AG 10 AMP NORMAL-BL
Fuse, 5 X 20 mm, 5 A, 250 V
Always disconnect power cord from the module before replacing a fuse.
Set voltage properly.
Use only the correct fuses.
Maintenance and Repair
Checking and/or Changing Power Fuses (F1)
8
99