Editing the Flow Rate
1 Press the RIGHT or LEFT ARROW to move the cursor on the FLOW field.
The values used in the following procedure assume that a 200 mL/min pump head is in use.
N O T E
• For 500 mL/min pump heads multiply the flow values by 2.5.
• For 800 mL/min pump heads multiply the flow values by 4.
• For 3.2 L/min pump heads multiply the flow values by 16.
2 To set the flow rate, type in the desired value and confirm with ENTER. For
Editing the Ramp Time
1 Press the RIGHT or LEFT ARROW to move the cursor on the ramp TIME field.
2 To set the ramp time, type in a decimal value and confirm with ENTER.
Start Pumping
1 Press the RIGHT or LEFT ARROW to move the cursor on the ramp TIME field.
2 To set the ramp TIME to 30 s, type in .5 and confirm with ENTER.
3 Press the RIGHT ARROW to move the cursor on the FLOW field.
4 To set the flow rate to 25 mL/min, type in the 25 and confirm with ENTER.
5 Press RUN to start the pump.
SD-1 Solvent Delivery - User Manual
example set the flow rate to 40 mL/min, type in 40 and confirm with ENTER.
You will hear the pump begin its 30 s ramp, and on the top line of the
display you can see flow and pressure values increasing. The elapsed ramp
time will also be shown.
Using the SD-1 Solvent Delivery Module
4
Initial Operation