Agilent Technologies SD-1 User Manual page 91

Isocratic solvent delivery module
Table of Contents

Advertisement

6 Hold the pump head and insert a 3/16 inch T-handle wrench into one of the
7 Tighten the other mounting screw until you feel resistance.
8 Tighten each mounting screw in turn a little at a time until both screws are
9 Replace the fittings at the outlet and inlet check valves.
10 Repeat the procedure for the other pump head.
SD-1 Solvent Delivery - User Manual
mounting screw holes and tighten the mounting screw until you feel some
resistance.
tight.
Maintenance and Repair
Replacing Seals
8
91