Site Requirements and Specifications

Performance Specifications

Performance Specifications
Table 7
Type
Display
Interface
CPU Board Voltages
RAM Backup Battery
Voltage
Table 8
Nominal flow
200 mL/min
500 mL/min
28
Performance specifications Agilent SD-1 Purification Solution
Specification
Backlit LCD with 4 lines, 160 characters
• 1 digital serial input/output channel (RS-422)
• 6 contact-closure relay outputs (1 A, 24 V)
• 5 high-speed CMOS contact-closure logic inputs
high: 3.5 – 6.5 V
low: –1.5 – 1 V1.5
max. input voltage 25 V
min. input voltage –20 V
min. on time for recognition 48.9 ms
• 1 programmable analog (D to A) output (0 – 10 V )
• 12 bit
• Absolute maximum error ±4.1 mV
• Test load impedance 100 kΩ
• 2 analog (A to D) inputs (–0.5 – 2.5 V )
• 60 Hz max. sample rate combined
• 5 V (4.75 – 5.25 V)
• –15 V
• 15 V
• 48 V
• 3.1 – 3.3 V
Pressure limits pump heads
Range
0.01 – 200 mL/min
0.02 – 500 mL/min
Maximum pressure
Maximum pressure
41.4 MPa
6000 psi
414 bar
17.2 MPa
2500 psi
172 bar