Must Stop Pump To Disengage Heads; Disengage Failed On Head 1; Heads Do Not Match - Agilent Technologies SD-1 User Manual

Isocratic solvent delivery module
Table of Contents

Advertisement

7
Error Information
Pump Error Messages

MUST STOP PUMP TO DISENGAGE HEADS

Probable cause
1

DISENGAGE FAILED ON HEAD 1

Probable cause
1
DISENGAGE FAILED ON HEAD 2
Probable cause
1

HEADS DO NOT MATCH

RIGHT HEAD: XXX mL/min, xxxxxx psi LEFT HEAD: XXX mL/min, xxxxxx
psi
Probable cause
1
74
To disengage the heads, the pump must be
stopped.
Disengage failed on head 1.
Disengage failed on head 2.
Heads don't match.
Suggested actions
Press STOP, or CLEAR to cancel.
Suggested actions
Press ENTER to continue.
Suggested actions
Press ENTER to continue.
Suggested actions
Press ENTER to continue.
SD-1 Solvent Delivery - User Manual