Troubleshooting
If You Encounter a Problem

Running Diagnostic Tests

Diagnostic tests, referred to as loop back tests, are provided to help isolate a problem. Perform the tests, in
the order listed, in the following flow chart.
Cycle the power on the N5102A digital module
Device Interface Test
N5102A Digital Bus Test
Digital Bus Cable Test
Signal Generator Digital Bus Test
Testing indicates a problem with the
signal generator. Send the signal
generator to Agilent for service. See
page 113
108
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Start here
page 109
Yes
Pass
No
page 110
Yes
Pass
No
page 111
Yes
Pass
No
page 11
Yes
Pass
No
for details.
End
The system is operating
correctly.
Send the N5102A
module to Agilent.
See
page 113
for
information on
returning the
module.
Replace the Digital Bus
cable. See
page 112
for
information on
replaceable parts.