4
Using the SD-1 Solvent Delivery Module
Operation
3 Replace and tighten the mounting screws in each pump head with a 3/16
4 Replace and tighten the mounting screws in the inlet tee with a 5/32 inch
5 Replace and tighten the mounting screws in the outlet tee with a 9/64 inch
6 Replace the low pressure inlet fitting in the inlet tee with the ½ inch
7 Connect the inlet tubing (tee to inlet check valves).
8 Replace the high pressure outlet fitting in the outlet tee with the 5/16 inch
9 Press ENTER to confirm the mounting screws and plumbing are connected.
10 Connect the pressure sensor cables.
11 Press ENTER to complete the head replacement procedure.
50
inch wrench.
wrench.
wrench.
open-ended wrench and tighten ¼-turn past fingertight.
open-ended wrench and tighten ¼-turn past finger-tight.
The following message will appear: Plug in BOTH connectors. Press ENTER when
completed.
The following display will read (with appropriate values for the actual head
size):
Figure 30
Pump head replacement
SD-1 Solvent Delivery - User Manual