Priming the Pump Heads
All pump heads are self-priming.
Preparations
• The detector is bypassed.
• If a prime purge valve is connected, open the prime purge valve to relieve the pressure.
1 Tighten the outlet tee fitting ¼-turn past finger-tight.
2 To switch on the pump power, press the power switch.
3 If the Flow field is not flashing, press the RIGHT ARROW key to get the cursor
4 To prime the 200 mL/min pump head, type in 150 and confirm with ENTER.
5 Start the pump by pressing ENTER.
6 Press STOP to stop the pump.
7 Switch off the SD-1 by pressing the power switch.
SD-1 Solvent Delivery - User Manual
After a few moments the Flow display is displayed:
to the flow field.
OR
To prime the 500 mL/min pump head, type in 400 and confirm with ENTER.
OR
To prime the 800 mL/min pump head, type in 600 and confirm with ENTER.
OR
To prime the 3.2 L/min pump head, type in 2000 and confirm with ENTER.
The pump will start to run and solvent will be drawn into the solvent inlet
line. After the inlet line is full, let the pump continue to run for a few
moments until the pressure display is steady.
The SD-1 pump is now primed.
Using the SD-1 Solvent Delivery Module
4
Operation
51