8 To change the pressure units, move the cursor to the UNITS field by pressing
9 To move the cursor to the ANALOG PRESSURE MONITOR and to reach the
Removing the Pump Head
When
The pumps are shipped without the pump head installed. You will have to install the pump head
before beginning to run. You can also change pump heads at any time.
Tools required
Description
Wrench, open-ended, 5/16 inch
Wrench, open-ended, ½ inch
Wrench, 9/64 inch hexagonal, T-handle
Wrench, 5/32 inch hexagonal, T-handle
Wrench, 3/16 inch hexagonal, T-handle
Preparing the removal in the menu
1 To stop the module, press the STOP key.
2 To release pressure from the HPLC system, open the drain valve.
SD-1 Solvent Delivery - User Manual
If the system develops a leak or the solvent reservoir runs dry, the pumps
will stop. It is good safety practice to use this feature.
the RIGHT ARROW key.
a Scroll through the units with using the DOWN ARROW key. The units are:
psi, bar, and MPa.
hidden part of the display, press the RIGHT ARROW key.
The ANALOG PRESSURE MONITOR allows to send an analog pressure signal to
a suitable device. (Not available with all instrument control software
programs to control the pumping system.)
Do not switch off the Agilent SD-1; power must be on for the heads to be
removed.