KROHNE OPTIWAVE 7300 C Handbook page 57

Non-contact radar fmcw level meter
Hide thumbs Also See for OPTIWAVE 7300 C:
Table of Contents

Advertisement

OPTIWAVE 7300 C
Menu No. Step
A.1.5
01/2008 • 4000172401 - HB OPTIWAVE 7300 R03 en
Function
outputs
1
output 1: function
2
output 1: 4 mA
setting
3
output 1: 20 mA
setting
5
output 1: output
range
6
output 1: error
handling
7
output 1: HART®
address
8
output 2: function
3
9
output 2: 4 mA
setting
3
10
output 2: 20 mA
setting
3
11
output 2: output
range
3
12
output 2: error
handling
3
13
settings summary
save / cancel
Function description
Follow this procedure to
describe the output
characteristics.
Select an output
function to scale the
current values . This is
not displayed in the
operator mode.
This assigns a
measurement value to
4 mA.
This assigns a
measurement value to
20 mA.
This sets the effective
range of output 1 with or
without over-run.
This sets the behaviour
of current output 1 if an
error occurs.
Any HART® address
greater than 0 will
activate HART®
multidrop mode.
Current output stays
constant at 4 mA.
Select an output
function to scale the
current values . This is
not displayed in the
operator mode.
This assigns a
measurement value to
4 mA.
This assigns a
measurement value to
20 mA.
This sets the effective
range of output 1 with or
without over-run.
This sets the behaviour
of current output 1 if an
error occurs.
www.krohne.com
OPERATION
Selection list
Default
-
1 solid/liquid: level,
level
distance, volume
(mass), ullage
volume; 2+ liquids:
interface level,
interface distance,
interface volume,
layer, layer volume
min.-max: 0...20 m /
0 m / 0 ft
0...65.62 ft
min.-max: 0...90 m /
depends
0...295.29 ft
on the
output
function
3.8...20.5 mA
4...20 mA
(NAMUR), 4...20 mA
3.6 mA, 22 mA, Hold
22 mA
min.-max: 0...15
0
1 liquid: level,
interface
distance, volume
level
(mass), ullage
volume; 2+ liquids:
interface level,
interface distance,
interface volume,
layer, layer volume
min.-max: 0...20 m /
0 m / 0 ft
0...65.62 ft
min.-max: 0...90 m /
depends
0...295.29 ft
on the
output
function
3.8...20.5 mA
4...20 mA
(NAMUR), 4...20 mA
3.6 mA, 22 mA, Hold
22 mA
read only
save, cancel
save
2
6
57

Advertisement

Table of Contents
loading

Table of Contents