KROHNE OPTIWAVE 7300 C Handbook page 109

Non-contact radar fmcw level meter
Hide thumbs Also See for OPTIWAVE 7300 C:
Table of Contents

Advertisement

OPTIWAVE 7300 C
Equipment needed (not supplied):
3 mm Allen wrench (for step 1).
2.5 mm Allen wrench (for step 5).
TORX T10 (for step 6).
How to remove the terminal module
Loosen the cover stop screw on the terminal compartment cover.
1
Remove the terminal compartment cover.
2
Remove the plastic terminal cover.
3
Disconnect the wire connector from the back end electronics compartment.
4
5
Loosen the 2 screws shown in the illustration.
Loosen the screw for the earth terminal connection.
6
7
Remove the terminal block from the housing. Send it to an authorized agent for maintenance.
How to attach the terminal module
Put the terminal block into the housing.
Attach the earth terminal connection to the housing with the screw supplied.
Attach the terminal block to the housing with 2 screws supplied.
Connect the wires from the electronics compartment.
Attach the plastic terminal cover.
Attach the terminal compartment cover.
Tighten the cover stop screw.
01/2008 • 4000172401 - HB OPTIWAVE 7300 R03 en
www.krohne.com
SERVICE
9
109