Remote Start - Fuji Electric ZPAJ Instruction Manual

Ndir type infrared gas analyzar
Table of Contents

Advertisement

Example
Start Time
Cycle
Flow Time
ON/OFF
In case where auto calibration is carried out at the above setting.
Sunday
12:00
Cycle
Zero
Ch1
calibration
Span
(Ch1    4)
calibration
350sec
350sec
Flow time
(An example of "Ch1: through Ch4: enable", as given in Section 6.2.4 "Setting of auto
calibration components/range")
Setting range
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7 days)
Flow time
: 60 to 900 sec
Caution
When an auto calibration starts, the measurement screen appears automatically.
During auto calibration, any key operation is not permitted other than operations such
as key lock ON/OFF and "Forced stop of auto calibration" (see Section 6.4.2.2).
When the key lock is set at ON, even the "Forced stop of auto calibration" cannot be
performed. To cancel auto zero calibration forcedly, set the key lock to OFF and then
execute "Forced stop of auto calibration".
Turn on the power again after it is turned off (including the case of power failure) at
the time set as the next start time in auto calibration, and then repeat it in the set cycle.
When the hold setting is set to ON, the hold time of auto calibration contact and
measurement value output signal are extended after calibration for gas replacement time.

Remote start

Remote start input
SUN
12:00
1
day
Zero
350 sec
Ch1 Span
350 sec
Ch2 Span
350 sec
Ch3 Span
350 sec
Ch4 Span
300 sec
EX. time
300 sec
ON
Monday
12:00
Ch2
Ch3
Ch4
Span
Span
Span
calibration
calibration
calibration
350sec
350sec
300sec
(initial value 300 sec)
Tuesday
12:00
Replace-
ment time
300sec
With input (hold at least 1.5 sec.)
Without input
42
: Auto calibration
INZ-TN2ZPAJ-E