Remote Start - Fuji Electric ZPG Instruction Manual

Ndir type infrared gas analyzer
Hide thumbs Also See for ZPG:
Table of Contents

Advertisement

cases.
Example
Start Time
Cycle
Flow Time
ON/OFF
In case where auto calibration is carried out at the above setting.
Sunday
12:00
Cycle
Ch1
Zero
Span
calibration
calibration
350sec
350sec
Flow time
(An example of "Ch1 and Ch2: enable", as given in Section 6.2.4 "Auto Calibration Components/range")
Setting range
Cycle
: 1 to 99 hours or 1 to 40 days (initial value 7 days)
Flow time
: 60 to 900 sec
CAUTION
When an auto calibration starts, the measurement screen appears automatically.
During auto calibration, any key operation is not permitted other than operations such as key lock
ON/OFF and "Forced stop of auto calibration" (see Section 6.4.2.2). When the key lock is set at
ON, even the "Forced stop of auto calibration" cannot be performed. To cancel auto calibration
forcedly, set the key lock to OFF and then execute "Forced stop of auto calibration".
Turn on the power again after it is turned off (including the case of power failure) at the time set
as the next start time in auto calibration, and then repeat it in the set cycle.
When the hold setting is set to ON, the hold time of auto calibration contact and measurement
value output signal are extended after calibration for gas replacement time.

Remote start

Remote start input
12:00
SUN
day
1
350 sec
Zero
350 sec
Ch1 Span
350 sec
Ch2 Span
300 sec
EX. time
ON
Monday
12:00
Ch2
Replace-
Span
ment time
calibration
350sec
300sec
(initial value 300 sec)
Tuesday
12:00
With input (hold at least 1.5 sec.)
Without input
52
: Auto calibration
INZ-TN2ZPG-E

Advertisement

Table of Contents
loading

This manual is also suitable for:

ZpgaZpgdZpg3eZrg3

Table of Contents