User Guide
Appendix B
The specifications provided in this chapter apply to the AutoPulse Resuscitation System Model 100.
B.1
Patient Parameters
The AutoPulse System is designed for adults with weight of no more than 300 lbs. (136 kg) with chest
circumference of 29.9 to 51.2 in. (76 to 130 cm) and chest width of 9.8 to 15 in. (25 to 38 cm).
Warning: The AutoPulse System is intended for use on adults, 18 years of age or older.
B.2
LifeBand
The latex-free LifeBand is for single-use only. The LifeBand consists of a cover plate and two bands
integrated with a patient liner and compression pads with a Velcro
B.3
Operating Parameters
Table B-1
Operating Parameters
Category
Chest displacement
Physiological duty cycle
Compression rate
Compression modes
(user selectable)
Ventilation pause
(30:2 and 15:2 mode)
B.4
Platform Physical
Table B-2
Physical Specifications
Category
Manufacturer
Size (L×W×H)
Weight (excluding Battery)
Display
Technical Specifications
Specifications
Equal to 20% reduction in anterior-posterior chest depth.
50 ± 5%.
80 ± 5 compressions per minute.
•
30:2 (30 compressions with two 1.5 second ventilation pauses)
•
15:2 (15 compressions with two 1.5 second ventilation pauses)
•
Continuous compressions
Two pauses of 1.5 seconds.
Specifications
ZOLL Circulation, Inc.
32.5 in. by 17.6 in. by 3.0 in. (82.6 cm by 44.7 cm by 7.6 cm).
20.5 lbs. (9.3 kg).
Dot matrix liquid crystal display (LCD), actively backlit, adjustable contrast.
®
fastener.
Page B-1