Mycom S CLM 153
Error no.
NAMUR
Error message
class
E038
Mainte-
Delta limit exceeded
nance
E046
Failure
Installation factor CH 1exceeded
E047
Failure
Installation factor CH1 undershot
E048
Mainte-
Installation factor CH 2 exceeded
nance
E049
Mainte-
Installation factor CH 2 undershot
nance
E053
Failure
Actuator error
E054
Mainte-
Dosing time monitor
nance
E055
Failure
Display / measuring range of
main parameter 1 undershot
E056
Failure
Display / measuring range of
main parameter 2 undershot
E057
Failure
Display / measuring range of
main parameter 1 exceeded
E058
Failure
Display / measuring range of
main parameter 2 exceeded
E059
Failure
Temperature measuring range 1
undershot
E060
Failure
Temperature measuring range 2
undershot
E061
Failure
Temperature measuring range 1
exceeded
E062
Failure
Temperature measuring range 2
exceeded
E063
Failure
Current range current output 1
undershot
E064
Failure
Current range current output 1
exceeded
E065
Failure
Current range current output 2
undershot
E066
Failure
Current range current output 2
exceeded
E067
Mainte-
Setpoint controller / LS 1
nance
exceeded
E068
Mainte-
Setpoint controller / LS 2
nance
exceeded
E069
Mainte-
Setpoint controller / LS 3
nance
exceeded
E070
Mainte-
Setpoint controller / LS 4
nance
exceeded
E071
Mainte-
Setpoint controller / LS 5
nance
exceeded
Endress+Hauser
Possible causes / measures
Check single measured values (input
1 / input 2) for plausibility Measuring
can still continue until failure error
E019 occurs.
Check pipe diameter, clean sensor
and carry out calibration again (only
for inductive).
Feedback to controller defective
Immerse sensor in conductive
medium.
For inductive: Carry out airset.
As long as the sensor is out of the
medium during Airset, an error code
referring to measuring range violation
can be ignored.
Temperature sensor defective;
Sensor line interrupted or short-
circuited;
Incorrect sensor type selected
Simulation s. chap. 7.1.3 /
chap. 7.1.4.
Measured value outside specified
current range:
Check measured value for plausibility,
if necessary adjust current output
assignment 0/4 mA and/or 20 mA.
Dosing devices defective;
chemical supply empty;
measured value incorrect
check
for plausibility and function;
Incorrect control direction set;
incorrect contact assigned;
Incorrect control function assigned
8 Troubleshooting
Alarm
Error
contact
current
Fact.
User
Fact.
User
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
yes
no
Automatic
cleaning
start
Fact.
User
no
no
no
no
no
no
no
no
no
no
89