8 Troubleshooting
⇒
!
Error no.
NAMUR
Error message
class
E001
Failure
Memory error
E002
Failure
User data invalid
E003
Failure
No valid configuration
E004
Failure
Erro invalid hardware or software
E006
Failure
Transmitter 2 error
E007
Failure
Transmitter 1 error
E008
Failure
Sensor error 1
E009
Failure
Sensor error 2
E010
Failure
Temperature sensor 1 defective
E011
Failure
Temperature sensor 2 defective
E019
Failure
Delta exceeded failure
E025
Failure
Limit Airset Offset CH 1 exceeded Carry out airset again (only inductive;
E026
Failure
Limit Airset Offset CH 2 exceeded
E034
Failure
Cell constant sensor 1 exceeded
E035
Failure
Cell constant sensor 1 undershot
E036
Failure
Cell constant sensor 2 exceeded
E037
Failure
Cell constant sensor 2 undershot
88
8.1.1
Error number list: Troubleshooting and configuration
In the following error list, you can find a description of all the error numbers occurring.
For each error number there is also data on whether the factory setting (= Fact.)
of this error triggers
• an alarm,
• an error current or
• cleaning.
To enter the error list, proceed as follows:
Note!
• Please process the errors as shown in Field H5 (Alarm menu) on page 45.
• The second column shows the assignment acc. to NAMUR work sheet NA64 (failure,
maintenance, function control).
Possible causes / measures
Switch instrument off and back on.
Corrective maintenance in the factory,
if necessary.
The new software cannot recognise
the module.
Test with new transmitter.
Check sensor and sensor connection
(chap. 7.1.5 / chap. 7.1.6 or by E+H
service)
Check temperature sensor and
connections; if necessary, check
transmitter with temperature
simulator.
Check single measured values
(input 1 / input 2) for plausibility
at air) or replace sensor. Clean and
dry sensor before airset.
Clean sensor and recalibrate; if
necessary, check sensor, line and
connections.
⇒
Alarm
contact
Fact.
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
yes
Mycom S CLM 153
Error
Automatic
current
cleaning
start
User
Fact.
User
Fact.
no
–
no
–
no
–
no
–
no
no
no
no
no
no
no
no
no
–
no
–
no
–
no
–
no
–
Endress+Hauser
User
–
–
–
–
–
–
Need help?
Do you have a question about the Mycom S CLM 153 and is the answer not in the manual?