Set up 1 – Emergency switching
If the hardware (e.g. sensor or transmitter) is defective, you can change the function of
the transmitter using the emergency switching. You can set the sensor of one circuit to
the transmitter of the other circuit.
To enter the menu, proceed as follows:
CODE
CHOICE
(default = bold)
N1
Caution! Being switched to
one-circuit measurement.
N2
Switching off
Sensor 1 --> input 2
Sensor 2 --> input 1
INFO
Linear
Select current output 1 / 2:
off
Alarm for characteristic value
100
(only related circuits)
200
Monitoring of measured value difference for two-circuit
measurement. Entry of the maximum permitted
difference at which the maintenance or failure alarm
should be triggered.
⇒
INFO
Note in display
Emergency switching
Sensor 1 is set to input 2 or vice versa.
The settings you made for the circuits are kept.The
circuit data apply after switching, with the exception of
the sensor-specific data.