Summary of Contents for Endress+Hauser Liquiline CM42B
Page 1
Products Solutions Services BA02380C/07/EN/01.25-00 71695529 2025-01-01 Operating Instructions Liquiline CM42B Two-wire transmitter Field device Measurement with digital or analog sensors...
About this document Liquiline CM42B About this document Warnings Structure of information Meaning This symbol alerts you to a dangerous situation. DANGER Failure to avoid the dangerous situation will result in a fatal or serious Causes (/consequences) injury. If necessary, Consequences of non-compliance (if applicable) ‣...
Liquiline CM42B Basic safety instructions Basic safety instructions Requirements for personnel • Installation, commissioning, operation and maintenance of the measuring system may be carried out only by specially trained technical personnel. • The technical personnel must be authorized by the plant operator to carry out the specified activities.
Basic safety instructions Liquiline CM42B Operational safety Before commissioning the entire measuring point: 1. Verify that all connections are correct. 2. Ensure that electrical cables and hose connections are undamaged. Procedure for damaged products: 1. Do not operate damaged products, and protect them against unintentional operation.
Liquiline CM42B Product description 3.1.3 Measuring parameters Depending on the order, the transmitter is designed for digital Memosens sensors or for analog sensors. A transmitter for analog sensors can be reconfigured to Memosens. This requires an activation code and the analog input module must be removed.
• Safety information and warnings • Ex markings • Certification information • Warnings ‣ Compare the information on the nameplate with the order. 4.2.2 Identifying the product Manufacturer address Endress+Hauser Conducta GmbH+Co. KG Dieselstraße 24 70839 Gerlingen Germany Product page www.endress.com/CM42B Endress+Hauser...
A new window opens. Here you fill information pertaining to your device, including the product documentation. Scope of delivery The scope of delivery includes: • Liquiline CM42B • Cable glands depending on order • Field device mounting plate • Brief Operating Instructions •...
Mounting Liquiline CM42B Mounting the device 5.2.1 Wall mounting 80 (3.15) ≥120 (4.72) ≥200 (7.87) A0053942 6 Mounting clearances in mm (in) A0053945 7 Wall mounting Wall Four drill holes Mounting plate Screws (not included in the scope of delivery) The size of the drill holes depends on the mounting material used.
Liquiline CM42B Mounting A0053943 8 Mounting plate mounted on wall A0053944 9 Attach the device and click it into place 1. Place the device on the mounting plate. 2. Slide the device downwards in the guide on the mounting rail until it clicks into place.
Page 16
Mounting Liquiline CM42B Weather protection cover (optional) Spring washers and nuts (post mounting kit) Post mounting plate (post mounting kit) Pipe or post (circular/square) Spring washers and nuts (post mounting kit) Mounting plate Pipe clamps (post mounting kit) Screws (post mounting kit) A0053916 ...
Liquiline CM42B Mounting 5.2.3 Rail mounting You require the post mounting kit (optional) to mount the unit on a pipe, post or railing (square or circular, clamping range 20 to 61 mm (0.79 to 2.40")). A0012668 13 Rail mounting...
Mounting Liquiline CM42B 5.2.4 Disassembly (for conversion, cleaning, etc.) CAUTION Risk of injury and damage to the device if the device is dropped ‣ When pushing the housing out of the holder, secure the housing to prevent it from falling.
Installation in hazardous areas Installation in hazardous area Ex ia Ga 4 … 20 mA A0056644 Hazardous area version of Liquiline CM42B Control station 4 to 20 mA signal line/optional HART Ex ia active barrier Supply and signal circuit Ex ia (4 to 20 mA)
Electrical connection Liquiline CM42B Connecting the device 6.2.1 Opening the housing NOTICE Cordless screwdriver, screw driller, pointed or sharp tools The use of a cordless screwdriver or screw driller can cause damage to the threads and impair the leak-tightness of the housing. If unsuitable tools are used, they can scratch the housing or damage the seal, and thus have a negative impact on the leak-tightness of the housing.
Liquiline CM42B Electrical connection 6.2.2 Connecting the cable shield The descriptions of each of the connections specify which cables must be shielded. Only use terminated original cables where possible. Clamping range of grounding clamps: 4 to 11 mm (0.16 to 0.43 in) Cable sample (does not necessarily correspond to the original cable supplied) ...
Electrical connection Liquiline CM42B 8. Connect cable cores as per the wiring diagram. 9. Tighten the cable gland with the required torque. 6.2.3 Cable terminals Press the screwdriver against the clip (opens the terminal). Insert the cable until the limit stop.
Page 23
Liquiline CM42B Electrical connection A0055833 1. Remove the sealing plug. 2. Screw in the cable gland. Tightening torque 2.5 to 3 Nm. Cable glands with G1/2 thread or NPT1/2 thread The cable glands and adapters are included in the delivery in accordance with the order.
Page 24
Electrical connection Liquiline CM42B A0057259 Tighten the cable gland again after the cable has been fed through. Make sure that the sealing insert (1) does not protrude from the pressure screw (2). Feed through only one cable per cable gland.
Page 25
Liquiline CM42B Electrical connection A0057685 Example: Three conduit adapters mounted at positions 2, 3 and 4 Example: Two conduit adapters mounted at positions 2 and 4 A0057686 Remove the sealing plug. 2. Remove the screw, securing disk and retaining plate from the potential equalization connection.
Page 26
Electrical connection Liquiline CM42B A0057687 Screw in the conduit adapter. Tightening torque 2.5 to 3 Nm. A0057690 Fit the conduit adapter support on the adapters or sealing plugs. Where necessary, align the adapters or sealing plugs by turning them. 5. Bolt the conduit adapter support to the potential equalization connection using the screw and securing disk.
Liquiline CM42B Electrical connection 6.2.6 Connecting the potential equalization Connecting the potential equalization – Installation without a conduit ‣ A0055870 21 Potential equalization connection Attach the potential equalization connection of the housing to the earth or to the potential equalization system with a separate line. Cable cross-section max.
Page 28
Electrical connection Liquiline CM42B Connecting the potential equalization for conduit installation ‣ A0057719 22 Conduit adapter support Connections for potential equalization A0057705 23 Potential equalization connection for conduit mounting For conduit mounting, connect the ground cable to a potential equalization connection of the conduit adapter support.
Page 29
Liquiline CM42B Electrical connection 4...20 mA 4...20 mA 4...20 mA/HART 4...20 mA/HART A0054900 24 Connection of 1 current output 4...20 mA/ HART – A0054914 25 Wiring diagram: 1 current output 4...20 mA /(HART) 4...20 mA/(HART) + 4...20 mA + 4...20 mA...
Electrical connection Liquiline CM42B 4...20 mA 4...20 mA 4...20 mA/(HART) 4...20 mA/(HART) A0054902 27 Connection of 2 current outputs via 2 cables 4...20 mA/ HART 4 ... 20 mA – – A0054915 28 Wiring diagram: 2 current outputs 6.2.8...
Page 31
Liquiline CM42B Electrical connection Explanation of color codes in the following illustrations: Color code Meaning Black Brown Blue Green Orange Yellow Violet White Transparent Braided shield/silver Memosens sensors Connecting sensors with Memosens plug-in head (via Memosens cable) and sensors with a fixed cable and Memosens protocol ‣...
Page 33
Liquiline CM42B Electrical connection Analog conductivity sensors (conductive) ‣ 11 13 12 20 19 Cond. C 11 1312 20 19 d.n.c. d.n.c. Sensor d.n.c. Sensor d.n.c. A0055786 33 Device view Device d.n.c. Cable d.n.c. Sensor A0055795 34 Wiring diagram Connect the sensor as shown in the illustration.
Page 34
Electrical connection Liquiline CM42B 1. Completely remove the semi-conductive polymer layer (3) up to the end of the shield. 2. Ensure that the inner insulation (4) of the coaxial cable is not in contact with other components. Ensure there is an air gap around all components; otherwise, measurement errors may occur.
Electrical connection Liquiline CM42B • Unsuitable cable diameters are used for the cable glands • The housing cover is not properly secured (danger of moisture entering due to inadequate sealing) • Cables/cable ends are loose or insufficiently tightened • Cable shields not grounded using grounding clamp in accordance with the instructions •...
Liquiline CM42B Operation options Operation options Overview of operation options Operation and settings via: • Operating elements on the device • SmartBlue app (does not support the full range of functions) • Control station (via HART) Access to operating menu via onsite display 7.2.1...
Operation options Liquiline CM42B 7.2.3 Structure of the display EH_CM42B_0123456789 14:00:20 2024-08-20 C 1: pH 7.00 Current Output 1 12.0 MENU DIAG A0056328 45 Structure of the display: Start screen (device with one current output) Device name or menu path...
Liquiline CM42B Operation options 7.2.4 Navigating through the display Measured values EH_CM42B_0123456789 EH_CM42B_0123456789 14:00:20 2024-08-20 14:00:20 2024-08-20 C 1: pH C 1: pH 7.00 7.00 Current Output 1 Current Output 1 12.0 12.0 MENU DIAG EH_CM42B_0123456789 EH_CM42B_0123456789 14:00:20 2024-08-20 14:00:20...
Page 44
Operation options Liquiline CM42B Current output EH_CM42B_0123456789 EH_CM42B_0123456789 14:00:20 2024-08-20 14:00:20 2024-08-20 C 1: pH C 1: pH 7.00 7.00 Current Output 1 Current Output 1 12.0 12.0 MENU DIAG EH_CM42B_0123456789 EH_CM42B_0123456789 14:00:20 2024-08-20 2024-08-20 Current output 1 C 1: pH 7.00...
Page 45
Liquiline CM42B Operation options 7.2.5 Operation concept menus EH_CM42B_0123456789 14:00:20 2024-08-20 Language English C 1: pH ► Guidance 7.00 ► Diagnostics ► Application Current Output 1 ► System 12.0 MENU DIAG Menu/Language Menu/Language English Language English ► Guidance Hrvatski Čeština ►...
Operation options Liquiline CM42B Access to the operating menu via the operating tool 7.3.1 Access to the operating menu via the SmartBlue app The SmartBlue app is available for download from the Google Play Store for Android devices and from the Apple App Store for iOS devices.
Page 47
Liquiline CM42B Operation options 7.3.3 Functions via the SmartBlue app The SmartBlue app supports the following functions: • Firmware update • User management • Export of information for the service Endress+Hauser...
System integration Liquiline CM42B System integration Integrating the measuring instrument into the system Interfaces for measured value transmission (depending on order): • 4 to 20 mA current output (passive) • HART 8.1.1 Current output Depending on the order, the device has 1 or 2 current outputs.
Page 49
Liquiline CM42B System integration Follow the steps below to integrate the device into the system for this purpose: 1. Connect the HART modem or HART handheld terminal to current output 1 (communication load 250–500 Ohm). 2. Establish a connection via the HART device.
Commissioning Liquiline CM42B Commissioning Preliminaries ‣ Connect the device. The device starts and displays the measured value. Bluetooth® must be enabled on the mobile device for operation via the SmartBlue app. Post-installation and function check Incorrect connection, incorrect supply voltage Safety risks for staff and device malfunctions! ‣...
Page 51
Liquiline CM42B Commissioning Initial access data: • Username: admin • Password: serial number of the device It is advisable to change the username and password after logging in for the first time. You can drag additional information (e.g. main menu) onto the screen by swiping across the screen.
Commissioning Liquiline CM42B Time and date ‣ Configure the time and date under the following path: Menu/System/Date and Time When using the SmartBlue app, the date and time can also be automatically transferred from the mobile device. Configuring the operating language ‣...
Liquiline CM42B Operation Operation 10.1 Reading the measured values See → 41 10.2 Adapting the measuring instrument to the process conditions 10.2.1 Calibrating the sensor Different calibration methods are available depending on the measuring parameter and connected sensor. 1. Navigate to the path: Menu/Guidance/Calibration or press the CAL softkey.
Operation Liquiline CM42B Calibration methods Different calibration methods are available depending on the measuring parameter and the sensor used. Configuration options: Select the calibration methods that are displayed under Menu/Guidance/Calibration ‣ Navigate to the path: Menu/Application/Sensor/Calibration settings/Calibration methods Further calibration settings Further calibration settings are available depending on the measuring parameter and connected sensor.
Diagnosis and troubleshooting Liquiline CM42B Diagnosis and troubleshooting 11.1 General troubleshooting The transmitter continuously monitors its functions itself. If a diagnostic message occurs, the display alternates between the diagnostic message and the measured value in the measuring mode . In the DIAG/Diagnostics list menu, you can find more detailed information on the current diagnostic messages displayed.
Liquiline CM42B Diagnosis and troubleshooting 11.5 Adapting the diagnostic information Configuration options: • Specify the error category for the diagnostic messages as per NAMUR NE 107 (F, M, C, • Specify the diagnostic behavior for the diagnostic messages. ‣ Navigate to the path: Menu/Diagnostics/Diagnostics settings 11.6...
Maintenance Liquiline CM42B Maintenance The maintenance of the measuring point comprises: • Calibration • Cleaning the transmitter, assembly and sensor • Checking cables and connections. WARNING Process pressure and temperature, contamination Risk of serious or fatal injury ‣ If the sensor has to be removed during maintenance work, avoid dangers posed by pressure, temperature and contamination.
Liquiline CM42B Repair Repair 13.1 General information ‣ Only use spare parts from Endress+Hauser to guarantee the safe and stable functioning of the device. Detailed information on the spare parts is available at: www.endress.com/device-viewer 13.2 Return The product must be returned if repairs or a factory calibration are required, or if the wrong product was ordered or delivered.
Page 60
Repair Liquiline CM42B BASE2-E Press together the securing clips of the base module and pull them out of the base module. Release the lithium button cell from the backplane and dispose of it according to local battery regulations. If required by the Directive 2012/19/EU on waste electrical and electronic equipment (WEEE), the product is marked with the depicted symbol in order to minimize the disposal of WEEE as unsorted municipal waste.
Liquiline CM42B Accessories Accessories The latest list of accessories, all compatible sensors and activation codes is provided on the product page: www.endress.com/CM42B Endress+Hauser...
• Cable length max. 100 m (330 ft) pH/ORP analog input Cable specifications Analog pH sensors and analog ORP sensors from Endress+Hauser • Recommended cable length max. 30 m (98 ft) • For cable types, see the documentation of the connected sensor Pfaudler electrodes type 03/04, type 18, type 40, pH Reiner Cable length max.
Liquiline CM42B Technical data Analog input of conductivity, measured conductively Cable specifications • Cable length max. 15 m (49.2 ft) • For cable types, see the documentation of the connected sensor Temperature sensors • Pt100 • Pt1000 15.2 Output Output signal...
Technical data Liquiline CM42B R [Ω] 1000 U [V] A0055514 Supply voltage [V] Load [Ω] Max. load with configured failure current 23 mA Max. load with configured failure current 21.5 mA Max. load with configured failure current < 3.6 mA Output span 3.6 to 23 mA...
Liquiline CM42B Technical data • For supply voltage, see characteristic curve • Max. supply voltage: 30 V DC U [V] 15.5 I [mA] A0055525 50 Min. supply voltage at the transmitter depending on the output current Supply voltage [V DC]...
Technical data Liquiline CM42B 15.6 Environment Ambient temperature Non-Ex version -30 to 70 °C (-20 to 160 °F) For Ex versions, please refer to the relevant safety instructions (XA) on the online product pages. Storage temperature -40 to +80 °C (-40 to 176 °F)