Application Errors - Endress+Hauser Fieldbus Levelflex M FMP43 Operating Instructions Manual

Guided level-radar
Hide thumbs Also See for Fieldbus Levelflex M FMP43:
Table of Contents

Advertisement

Troubleshooting
Error
Output
A warning or alarm
Depending on the configuration
has occurred.
Measured value
(00) is incorrect
No change of
measured value on
filling/emptying
E 641 (loss of
If the device is configured to HOLD in the event
echo) after
of echo loss the output is set to any value/
switching on the
current.
power supply
84
9.3

Application errors

100%
F m/ft
expected
actual
E m/ft
0%
L00-FMP4xxxx-19-00-00-en-019
100%
actual
expected
0%
L00-FMR2xxxx-19-00-00-en-014
Levelflex M FMP43 with FOUNDATION Fieldbus
Possible cause
See table of
error messages
(→ ä 82)
Measured distance
(008) OK?
t →
no
An interference echo
may have been
evaluated.
Interference echo from
tank internals, nozzle
or buildup on the
probe
t →
Noise level during the
initialization phase too
high.
Remedy
1. See table of error messages
(→ ä 82)
yes
1. Check empty calibr. (005) and
full calibr. (006)
2. Check linearization:
Level/ullage (040)
Max. scale (046)
Diameter vessel (047)
Check table
yes
1. Carry out tank mapping
Basic setup
1. Carry out tank mapping
Basic setup
2. If necessary, clean probe
3. If necessary, select better
mounting position
Repeat empty calibr. (005) once
more.
Caution!
Before confirmation change with
S
or
to the edit mode.
Endress+Hauser
O

Advertisement

Table of Contents
loading

Table of Contents