Envelope Curve With Vu331 - Endress+Hauser Fieldbus Levelflex M FMP43 Operating Instructions Manual

Guided level-radar
Hide thumbs Also See for Fieldbus Levelflex M FMP43:
Table of Contents

Advertisement

Commissioning
ENDRESS + HAUSER
+
E
ENDRESS + HAUSER
+
E
!
68
6.6

Envelope curve with VU331

After the basic setup, it is recommended to evaluate the measurement with the aid of the envelope
curve ("envelope curve" (0E)) function group)
6.6.1
Function "plot settings" (0E1)
Here you can select which information is shown on the display:
• Envelope curve
• Substracted signal
• Mapping
6.6.2
Function "recording curve" (0E2)
This function determines whether the envelope curve is read as
• Single curve or
• Cyclic
Note!
If the envelope curve mode is active on the display, the measured values are updated in a slower
cycle time. Thus, it is advisable to leave the envelope curve mode after the measuring point has been
optimized.
Levelflex M FMP43 with FOUNDATION Fieldbus
Endress+Hauser

Advertisement