Retest; Intensive Test; Test The Next Profile - Novatest ZBL-U5 Series Instruction Manual

Ultrasonic detectors
Table of Contents

Advertisement

Manual for ZBL-U5 series of Ultrasonic Detector

4.3.3 Retest

Retest refers to testing the part or all of the tested data once again.
During sampling, any unsatisfying data may be retested at any time.
1)Automatic pile test mode: In dynamic sampling, if transducer
is placed back and lifted again, retest will be implemented
automatically. New test data will cover the previous data
automatically in retest.
2)Manual pile test mode: In dynamic sampling, after have placed
the transducer at the height to be retested, press Retest to pop up a
dialogue box requiring input the starting point to be retested (the
default of start position is the depth of last test point). It cannot be
larger than the height of the first point in upward test or smaller than
the height of the first point in downward test; retest will start after has
been confirmed and current data will be covered.
Note: Retest method in manual mode is applicable to
automatic mode as well.

4.3.4 Intensive test

Intensive test refers to the test on suspicious area found in test
after test point spacing has been reduced. Press Interval button in main
interface to change test point spacing at any time.

4.3.5 Test the next profile

Place all transducers at proper position, click Profile button to
choose the profile to be tested and press Sample button to
automatically Start continuous sampling. The subsequent operation
is the same as section 4.3.1.
63

Advertisement

Table of Contents
loading

This manual is also suitable for:

Zbl-u5100Zbl-u5200Zbl-u5300

Table of Contents