Waveform Display And Operation - Novatest ZBL-U5 Series Instruction Manual

Ultrasonic detectors
Table of Contents

Advertisement

Manual for ZBL-U5 series of Ultrasonic Detector
keyboard is closed; if the input number is illegal or unreasonable,
relevant prompt will be shown on the bottom;<0}
5) After have pressed Cancel button, the input is invalid and
soft keyboard is closed.

3.5 Waveform display and operation

Single-channel waveform range is used to display the waveform
and acoustic parameters, etc. of current testing point as shown in
Figure 3.4. The bottom horizontal row of single-channel waveform
range for strength test, defect test and crack test displays acoustic
parameter and zero acoustic time value of head wave, the lower left
corner displays series number of test point, top left corner displays
delay time, as shown in Figure 3.4b.
The vertical progress bar at the right of single-channel waveform
means the gain; when gain is increasing, the progress bar ascends and
when gain is decreasing, the bar descends.
1)
Acoustic time value T: Transmission time of ultrasonic from
transmitting transducer to receiving transducer, the unit is μs,
this acoustic time value has deducted the system zero
Figure 3.4 Schematic graph of waveform