Manual for ZBL-U5 series of Ultrasonic Detector
2. Test of subsequent test points
After the first test point having been saved, move the two
transducers down to the next test point, press the Save button upon
properly adjustment of waveform, repeat such process again and again
until the completing of test on all test points.
3. Insert the Empty Test Points
Under the condition of dynamic sampling, press the lower right
corner of the waveform and pop up the waveform operation panel. As
shown in Figure 6.4. Press the Blank button to insert an empty test
point, data of which being displayed in "---"and not being calculated.
Figure 6.4 Operation panel for dynamic waveform
6.3.2

Stop sampling

During the sampling, it can be stopped by pressing the Stop button,
which will turn over to the Sample button, and the Search button will
return over to the Exit button at the moment, and the data will be
saved into the disk automatically.
6.3.3

Retest

Retest refers to testing the part or all of the tested data once again.
Please refer to chapter 3.7 for details. New test data will cover the
previous data automatically in retest.