Advanced Parameter - Novatest ZBL-U5 Series Instruction Manual

Ultrasonic detectors
Table of Contents

Advertisement

Manual for ZBL-U5 series of Ultrasonic Detector
5.
Pipe distance
It refers to the clear spacing of sounding pipe outer wall in mm.
System will generate corresponding sounding pipe spacing options
automatically according to the number of sounding pipe. Press edit
box behind corresponding profile name to set sounding pipe spacing
in pop-up numerical soft keyboard.
Note: Since sounding pipe spacing may affect data
calculation, please set it correctly.

4.2.2 Advanced parameter

Press Advanced button in parameter setup interface to pop up
dialogue box as shown in Figure 4.4. This dialogue box includes two
property pages of General parameter and Other parameters. Default
of each parameter is the value set at last time.
After have set all parameters, press OK button to save parameters
have been set, press Cancel button for not.
Press Reset button in advanced parameter setup interface to pop
up dialogue box inquiring "Whether restore parameter to factory
setting?"; press Yes button to restore all parameters to factory setting
and No for not.
Note: If tested data exist, Reset button is invalid.
4.2.2.1
General parameter
Press General label to switch to its property page for setting
parameters such as Instrument system delay, Technical specification
and Acoustic speed limit, etc.
53