2.8.5.
Calibrate sensors
MultiLab enables you to calibrate any of the linear sensors manually. This two point
calibration method sets both the gain (slope) and offset (intercept) of the sensor's
conversion function. The calibration procedure affects MultiLab readings only.
1. Tap the menu bar, tap Logger, then tap Calibrate sensors
2. Pick a sensor from the Choose sensor pick list
3. Tap OK
4. Enter a distinct real value in each of the Real Value edit boxes and
the corresponding measured values in each of the Measured Value
edit boxes (The measured values are the values displayed by
MultiLab when measuring the two real values
5. Tap OK
The calibrated sensor parameters will be saved in MultiLab.
To reset to the default calibration for any sensor, repeat steps 1 to 3 above and tap
Defaults.
Chapter 2 Working with Palm Handheld
45