To switch to 8 sensors mode tap the Auto detect sensors checkbox a second time.
2.8.3.

Power saving mode

When performing long experiments at low rates, of up to 1 per minute, TriLink
enables you to work in power saving mode. In this mode TriLink switches to standby
mode and 'wakes up' for brief periods of time only to execute data logging and then
returns to a standby.
This will enable TriLink to work continuously, without recharging the battery, for up to
100 hours instead of 5 hours in normal mode.
To enable TriLink's power saving mode:
1. Establish connection between TriLink and Palm
2. Tap Setup
3. Tap the Rate tab
4. Tap the Power saving mode checkbox to check it
5. Tap OK
Note: the Power saving mode check box is enabled only after selecting
recording rate that is equal to or less then 1 sample every minute.
2.8.4.

Triggering

To start the data recording only when a specific time or measurement condition has
been met:
1. Establish connection between TriLink and Palm
2. Tap Setup
3. Tap the Samples tab
4. Tap Triggering
on the main toolbar