Fourier TriLog User Manual page 35

Fourier trilog data loggers: user guide
Table of Contents

Advertisement

3. Specify units for the sensor (e.g. g)
4. Select Equation from the Calibrate by picker.
5. Referring to the documentation provided with the sensor, enter
values for the slope and the y-intercept of the calibration curve.
6. Tap Save
2.
Calibrating a Linear Sensor by Reference
To calibrate a sensor by reference, from the View Sensor form:
1. Tap New Calibration
2. Specify an operating range (Min to Max) for the sensor by entering
its minimum and maximum values
3. Specify units for the sensor (e.g. g)
4. Select Reference from the Calibrate by picker
5. Tap in the Reference Point picker to set the value of the first
reference point
6. Follow the instructions in the Set Calibration Point form to add
values for the first reference pair
7. Tap OK to accept values for the first reference pair
Chapter 2 Working with Palm Handheld
35